Signet 2724-2726 and 2734-2736 Series pH and ORP Electrodes
The 3-2726-LC pH electrodes can be used in
low conductivity water of less than 100 μS.
When used in the range of 20 to 100 μS,
the ow range must not exceed 1 m/s (3 ft/s) velocity.
When used in liquids of less than 20 μS,
the ow range must not exceed 150 ml/min;
the sensor should also be mounted in a well grounded cell.
Sensor Installation and Removal
In-Line Installation
Low Conductivity Installation
3-2750-X
or 3-2760-X1
In-Line
Preampli er
Electrode
Signet Installation Fitting DN15 to DN100 (½ in. to 4 in.)
Flexible
conduit
Lubricate O-rings
with non-petroleum
based, viscous
lubricant (grease)
compatible with
the system.
Flexible conduit
PTFE tape
Fitting (customer supplied)Electrode
3-2750-X or
3-2760-X
Submersible
Preampli er
Lubricate
O-rings with a
non-petroleum
based, viscous
lubricant (grease)
compatible with
the system.
See peel & stick warning label on back!
Attach near sensor
3-2700.615 (B-04/08)
1. Use appropriate eye, face, hand, body
and/or respiratory protection when using
chemicals or solvents
2. Prior to installation or removal:
a. Depressurize and vent system
b. Drain below sensor level
3. Confirm chemical compatibility before use
4. Do not exceed the maximum
temperature/pressure specifications
5. Do not alter product construction
If installing into a threaded connection:
6. Inspect threads to ensure integrity. Do not
install a sensor that has damaged threads
7. Apply PTFE tape to the process connection
threads in accordance with industry
practices
8. Hand tighten the sensor into the process
connection. DO NOT USE TOOLS
Failure to follow
these instructions may
result in personal injury
WARNING
SAFETY INSTRUCTIONS
Horizontal
Yes
Yes
Yes
Yes
Yes
Yes
1. The electrode can be
mounted at any angle.
2. Avoid air pockets and sediment.
3. The tting must place the
electrode in the ow but
must not bottom out in the pipe.
4. Select a Signet installation
tting for convenience.
5. Use the ¾ in. threads on the
electrode body to install the
electrode into reducing tee ttings.
6. Inspect threads to ensure integrity. Do not install an electrode
with damaged threads.
7. Apply PTFE tape to the process connection threads,
in accordance with industry standards.
8. Use piping installation hardware with smooth,
well- nished threads to facilitate the installation.
9. If necessary, the pipe should be plumbed with a depression
(trap) so liquid is maintained around the electrode tip.
10. Hand-tighten the electrode into the process connection.
Do not use any tools to install the electrode.
The use of wrenches, pliers or similar may over-stress
the sensor body and lead to breakage and subsequent
spillage of the process liquid.
CAUTION:
A broken sensor may be
ejected forcefully from the tting
and can cause severe injury.
11. The safety instructions have an
adhesive label and should be placed
near the sensor.
Do not use as a handle!
Do not use tools or lubricant
to install yellow retaining cap.
Do not overtighten.
PTFE tape
3-2700.615 (B-04/08)
1. Use appropriate eye, face, hand, body
and/or respiratory protection when using
chemicals or solvents
2. Prior to installation or removal:
a. Depressurize and vent system
b. Drain below sensor level
3. Confirm chemical compatibility before use
4. Do not exceed the maximum
temperature/pressure specifications
5. Do not alter product construction
If installing into a threaded connection:
6. Inspect threads to ensure integrity. Do not
install a sensor that has damaged threads
7. Apply PTFE tape to the process connection
threads in accordance with industry
practices
8. Hand tighten the sensor into the process
connection. DO NOT USE TOOLS
Failure to follow
these instructions may
result in personal injury
WARNING
SAFETY INSTRUCTIONS
3-2700.615 (B-04/08)
1. Use appropriate eye, face, hand, body
and/or respiratory protection when using
chemicals or solvents
2. Prior to installation or removal:
a. Depressurize and vent system
b. Drain below sensor level
3. Confirm chemical compatibility before use
4. Do not exceed the maximum
temperature/pressure specifications
5. Do not alter product construction
If installing into a threaded connection:
6. Inspect threads to ensure integrity. Do not
install a sensor that has damaged threads
7. Apply PTFE tape to the process connection
threads in accordance with industry
practices
8. Hand tighten the sensor into the process
connection. DO NOT USE TOOLS
Failure to follow
these instructions may
result in personal injury
WARNING
SAFETY INSTRUCTIONS
3-2700.615 (B-04/08)
1. Use appropriate eye, face, hand, body
and/or respiratory protection when using
chemicals or solvents
2. Prior to installation or removal:
a. Depressurize and vent system
b. Drain below sensor level
3. Confirm chemical compatibility before use
4. Do not exceed the maximum
temperature/pressure specifications
5. Do not alter product construction
If installing into a threaded connection:
6. Inspect threads to ensure integrity. Do not
install a sensor that has damaged threads
7. Apply PTFE tape to the process connection
threads in accordance with industry
practices
8. Hand tighten the sensor into the process
connection. DO NOT USE TOOLS
Failure to follow
these instructions may
result in personal injury
WARNING
SAFETY INSTRUCTIONS
Lubricate
O-rings with
non-petroleum based,
viscous lubricant
(grease) compatible
with the system.